diff options
Diffstat (limited to '174/CH5/EX5.5/example5_5.sce')
-rwxr-xr-x | 174/CH5/EX5.5/example5_5.sce | 54 |
1 files changed, 54 insertions, 0 deletions
diff --git a/174/CH5/EX5.5/example5_5.sce b/174/CH5/EX5.5/example5_5.sce new file mode 100755 index 000000000..235bad6dc --- /dev/null +++ b/174/CH5/EX5.5/example5_5.sce @@ -0,0 +1,54 @@ +// To balance the unbalanced bridge
+// Modern Electronic Instrumentation And Measurement Techniques
+// By Albert D. Helfrick, William D. Cooper
+// First Edition Second Impression, 2009
+// Dorling Kindersly Pvt. Ltd. India
+// Example 5-5 in Page 119
+
+
+clear; clc; close;
+
+// Given data
+Z_1 = -1000*%i;
+Z_2 = 500;
+Z_3 = 1000;
+Z_4 = 100+500*%i;
+
+// The balance is not possible with this condition as theta_1+theta_4 will be slightly negative than theta_2+theta3
+// Balance can be achieved by 2 methods:
+disp('First option is to modify Z_1 so that its phase angle is decreased to less than 90deg by placing a resistor in parallel with the capacitor.')
+// The resistance R_1 can be determined by the standard approach
+
+//Calculations
+Y_1 = Z_4/(Z_2*Z_3);
+//Also,
+// Y_1 = (1/R) + %i/1000;
+// equating both the equations and solving for R_1
+
+R_1 = 1/(Y_1-(%i/1000 ));
+printf("The value of the resistor R_1 in parallel with capacitor = %d ohm\n",R_1);
+
+// It should be noted that the addition of R_1 upsets the first balance condition as the magnitude of Z_1 is changed
+// Hence the variable R_3 should be adjusted to compensate this effect
+
+disp('The second option is to modify the phase angle of arm 2 or arm 3 by adding series capacitor');
+Z_3_1 = Z_1 *Z_4/Z_2;
+// substituting for the component values and solving for X_C yeilds
+
+X_C = abs(1000- Z_3_1)/-%i;
+printf("The value of the reactance of the capacitor used, X_C = %d ohm",imag(X_C));
+
+
+//In this case the magnitude of the Z_3 is increased so that the first balance condition is changed
+//A small adjustment of R_3 is necessary to restore balance
+
+//Result
+// First option is to modify Z_1 so that its phase angle is decreased to less than 90deg by placing a resistor in parallel with the capacitor.
+// The value of the resistor R_1 in parallel with capacitor = 5000 ohm
+
+// The second option is to modify the phase angle of arm 2 or arm 3 by adding series capacitor
+// The value of the reactance of the capacitor used, X_C = 200 ohm
+
+
+
+
|